Computer-aided detection (CAD) of solid pulmonary nodules in chest x-ray equivalent ultralow dose chest CT - first in-vivo results at dose levels of 0.13mSv

Wissenschaftlicher Artikel/Review - 11.10.2016

Bereiche
PubMed
DOI

Zitation
Messerli M, Kluckert J, Knitel M, Rengier F, Warschkow R, Alkadhi H, Leschka S, Wildermuth S, Bauer R. Computer-aided detection (CAD) of solid pulmonary nodules in chest x-ray equivalent ultralow dose chest CT - first in-vivo results at dose levels of 0.13mSv. Eur J Radiol 2016; 85:2217-2224.
Art
Wissenschaftlicher Artikel/Review (Englisch)
Zeitschrift
Eur J Radiol 2016; 85
Veröffentlichungsdatum
11.10.2016
eISSN (Online)
1872-7727
Seiten
2217-2224
Kurzbeschreibung/Zielsetzung

OBJECTIVES
To determine the value of computer-aided detection (CAD) for solid pulmonary nodules in ultralow radiation dose single-energy computed tomography (CT) of the chest using third-generation dual-source CT at 100kV and fixed tube current at 70 mAs with tin filtration.

METHODS
202 consecutive patients undergoing clinically indicated standard dose chest CT (1.8±0.7 mSv) were prospectively included and scanned with an additional ultralow dose CT (0.13±0.01 mSv) in the same session. Standard of reference (SOR) was established by consensus reading of standard dose CT by two radiologists. CAD was performed in standard dose and ultralow dose CT with two different reconstruction kernels. CAD detection rate of nodules was evaluated including subgroups of different nodule sizes (<5, 5-7, >7mm). Sensitivity was further analysed in multivariable mixed effects logistic regression.

RESULTS
The SOR included 279 solid nodules (mean diameter 4.3±3.4mm, range 1-24mm). There was no significant difference in per-nodule sensitivity of CAD in standard dose with 70% compared to 68% in ultralow dose CT both overall and in different size subgroups (all p>0.05). CAD led to a significant increase of sensitivity for both radiologists reading the ultralow dose CT scans (all p<0.001). In multivariable analysis, the use of CAD (p<0.001), and nodule size (p<0.0001) were independent predictors for nodule detection, but not BMI (p=0.933) and the use of contrast agents (p=0.176).

CONCLUSIONS
Computer-aided detection of solid pulmonary nodules using ultralow dose CT with chest X-ray equivalent radiation dose has similar sensitivities to those from standard dose CT. Adding CAD in ultralow dose CT significantly improves the sensitivity of radiologists.