Publikation
Determination of the source dwell position of an afterloading device with a detector array
Wissenschaftlicher Artikel/Review - 06.09.2011
Breithuth F, Schiefer Johann, Arn Markus, Peters Samuel, Seelentag Wolf
Bereiche
Schlagwörter (Tags)
DOI
Kontakt
Zitation
Art
Zeitschrift
Veröffentlichungsdatum
eISSN (Online)
Verlag
Kurzbeschreibung/Zielsetzung
It was the aim to develop and test a measurement technique for the source position of an afterloading system using an electronic two-dimensional detector array (2D array). A "GammaMed plus IX" high dose rate afterloading device (Varian Medical Systems, Palo Alto, USA), and a Seven29 2D detector array (PTW Freiburg GmbH, Freiburg, Germany) have been used. A hollow needle has been connected to the afterloading device. Its outer diameter is 3.0 mm and its length about 220 mm. A 14 mm thick Perspex slab was fixed in a reproducible position on the 2D array. Above the 14th detector row, a groove was cut in the slab which accommodates the hollow needle in a fixed position relative to the 2D array. In order to define the position of the source, the signals of the detectors in the 14th detector row have been evaluated. A theoretical curve depending on the source position and strength, has been fitted to the acquired detector signals. It has been shown that the reproducibility of the measurement technique - including the reproducibility of the source placement - was within 0.15 mm. This method can not only replace film measurements, it is also more exact and less time consuming.