In vitro testing of a funnel-tip catheter with different clot types to decrease clot migration in mechanical thrombectomy.

Wissenschaftlicher Artikel/Review - 01.09.2022

Bereiche
PubMed
DOI

Zitation
Payne E, Krost-Reuhl S, Heimann A, Keric N, Masomi-Bornwasser J, Gerber T, Seidman L, Kirschner S, Brockmann M, Tanyildizi Y. In vitro testing of a funnel-tip catheter with different clot types to decrease clot migration in mechanical thrombectomy. Interv Neuroradiol 2022; 29:637-647.
Art
Wissenschaftlicher Artikel/Review (Englisch)
Zeitschrift
Interv Neuroradiol 2022; 29
Veröffentlichungsdatum
01.09.2022
eISSN (Online)
2385-2011
Seiten
637-647
Kurzbeschreibung/Zielsetzung

Mechanical thrombectomy is the standard treatment for acute ischemic stroke in patients with large vessel occlusion and can be performed up to 24h after symptom onset. Despite high recanalization rates, embolism in new territories has been reported in 8.6% of the cases. Causes for this could be clot abruption during stent retrieval into the smaller opening of a standard distal access catheter, and antegrade blood flow via collaterals despite proximal balloon protection. A funnel-shaped tip with a larger internal diameter was developed to increase the rate of first-pass recanalization and to improve the safety and efficacy of mechanical thrombectomy.