Publikation
Results of the TLD intercomparison 2014
Wissenschaftlicher Artikel/Review - 01.01.2014
Schiefer Johann, Heinze Simon
Bereiche
Schlagwörter (Tags)
Link
Kontakt
Zitation
Art
Zeitschrift
Veröffentlichungsdatum
Verlag
Kurzbeschreibung/Zielsetzung
It was the aim of this year's SSRMP intercomparison to check the absolute dosimetry of photon beams in a solid water phantom. The reference measurements have been performed in cooperation with Anton Steiner, METAS. Altogether 28 institutions participated in the intercomparison. 120 beams have been checked.
All (120) measured Linac doses agree with the stated doses within 4%, which is considered as a satisfactory check. 118 checked beams (98.3%) coincide within 3%, 89.2% of all Dm/Ds values are within the [0.98, 1.02] interval, and 55% within the [0.99, 1.01] interval.