Publication

Autoadjusted versus fixed CPAP for obstructive sleep apnoea: a multicentre, randomised equivalence trial

Journal Paper/Review - Oct 5, 2017

Units
PubMed
Doi

Citation
Bloch K, Matthews F, Laube I, Imhof E, Turk A, Schoch O, Kohler M, Russi E, Senn O, Nussbaumer-Ochsner Y, Lo Cascio C, Latshang T, Furian M, Huber F, Thurnheer R. Autoadjusted versus fixed CPAP for obstructive sleep apnoea: a multicentre, randomised equivalence trial. Thorax 2017; 73:174-184.
Type
Journal Paper/Review (English)
Journal
Thorax 2017; 73
Publication Date
Oct 5, 2017
Issn Electronic
1468-3296
Pages
174-184
Brief description/objective

BACKGROUND
The obstructive sleep apnoea syndrome (OSAS) is conventionally treated by continuous positive airway pressure set at a fixed level (fCPAP). Automatic mask pressure adjustment (autoCPAP) is increasingly used during home therapy. We investigated whether autoCPAP is equivalent to fCPAP in improving sleepiness in patients with OSAS in the long-term.

METHODS
In this multicentre equivalence trial, 208 patients with OSAS, with median Epworth sleepiness score (ESS) 13, apnoea/hypopnoea index 48.4/hour, were randomised to treatment with autoCPAP (5-15 mbar) or fCPAP (pressure set at the 90th percentile applied by autoCPAP during 2-4 weeks adaptation). Coprimary outcomes were changes in subjective and objective sleepiness from baseline to 2 years after treatment. Equivalence ranges were ±2 points in ESS and ±3 min sleep resistance time evaluated by recording responses to light signals.

RESULTS
At 2 years, in the intention to treat analysis, the reduction in sleepiness versus pretreatment baseline was similar in patients using autoCPAP (n=113, mean ESS-change -6.3, 95% CI -7.1 to -5.5; sleep resistance time +8.3 min, +6.9 to +9.7) and fCPAP (n=95, mean ESS-change -6.2, 95% CI -7.0 to -5.3; sleep resistance time +6.3 min, +4.7 to +7.8). The 95% CI of difference in ESS-reduction between autoCPAP and fCPAP was -0.9 to +1.4 and the 95% CI of difference in increase in sleep resistance time was -2.6 to +1.0 min. Blood pressure reduction and OSAS-related costs were similar between groups.

CONCLUSIONS
AutoCPAP and fCPAP are equivalent within prespecified ranges in improving subjective and objective sleepiness in patients with OSAS over the course of 2 years. Costs of these treatments are similar.

TRIAL REGISTRATION NUMBER
ClinicalTrials.gov NCT00280800.