Publication

Accuracy of computer-assisted pedicle screw placement. An in vivo computed tomography analysis

Journal Paper/Review - Feb 15, 1997

Units
PubMed

Citation
Schwarzenbach O, Berlemann U, Jost B, Visarius H, Arm E, Langlotz F, Nolte L, Ozdoba C. Accuracy of computer-assisted pedicle screw placement. An in vivo computed tomography analysis. Spine 1997; 22:452-8.
Type
Journal Paper/Review (English)
Journal
Spine 1997; 22
Publication Date
Feb 15, 1997
Issn Print
0362-2436
Pages
452-8
Brief description/objective

STUDY DESIGN
A computer-assisted planning and visualization system (the Orthopaedic Surgery Planning System) was tested for pedicle screw insertion in vivo.

OBJECTIVES
To evaluate the system's applicability for regular intraoperative use and its accuracy for pedicle screw placement in vivo.

SUMMARY OF BACKGROUND DATA
Pedicle screw placement poses considerable anatomic and biomechanical risks. The reported rates of screw misplacement with conventional insertion techniques are unacceptably high. It previously has been shown in vitro that computer assistance offers the potential to decrease the number of screws perforating the pedicular cortex.

METHODS
The accuracy of 162 pedicle screws inserted with the Orthopaedic Surgery Planning System was assessed by means of postoperative computed tomography evaluation. Reconstructions of the horizontal, frontal, and sagittal planes were analyzed. Cortex perforations were graded in steps of 2 mm.

RESULTS
The cortex was perforated in 2.7% of pedicles. Complete preoperative computed tomography scanning of the levels to be operated on is essential to allow for a precise image reconstruction. Initial difficulties in applying the system contribute to the malplacements. A learning curve for general handling of the Orthopaedic Surgery Planning System was observed.

CONCLUSIONS
The system provides a safe and reproducible technique for pedicle screw insertion. Other applications in the field of spine surgery are under evaluation.