Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey EM-4000 and Nidek CEM-530.

Journal Paper/Review - Jul 9, 2024

Units
PubMed
Doi
Contact

Citation
Kecik M, Kropp M, Thumann G, Pajic B, Guber J, Guber I. Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey EM-4000 and Nidek CEM-530. Medicina (Kaunas) 2024; 60
Type
Journal Paper/Review (English)
Journal
Medicina (Kaunas) 2024; 60
Publication Date
Jul 9, 2024
Issn Electronic
1648-9144
Brief description/objective

: The purpose of this study was to compare two commercially available specular microscopes (Tomey EM-4000 and Nidek CEM-530) in a real-life clinical setting in terms of intra- and interdevice variability. The study was conducted on all patients seen in a clinical practice specializing in anterior segment pathologies, regardless of the purpose of their visit. : In total, 112 eyes of 56 patients (age 23-85 years old) were included in the study. Each eye was measured three times with each device (for a total of six measurements), and results for central corneal thickness (CCT) and corneal endothelial cell density (ECD) were recorded. The results were then evaluated with the D'Agostino-Pearson normality test and compared with a Wilcoxon signed-rank test, -test, ANOVA or Mann-Whitney test for intra- and interdevice variability. : Both specular microscopes produced very reliable reproducible intradevice results: The Tomey EM-4000 measured an ECD of 2390 ± 49.57 cells/mm (mean ± standard error of mean); the range was 799-3010 cells/mm. The determined CCT was 546 ± 5.104 µm (mean ± standard error of mean [SEM]); the range was 425-615 µm. The measurements with the Nidek CEM-530 revealed an ECD of 2417 ± 0.09 cells/mm (mean ± SEM); the range was 505-3461 cells/mm (mean ± SEM). The mean CCT detected was 546.3 ± 4.937 µm (mean ± SEM); the range was 431-621 µm. The interdevice differences were statistically significant for both parameters, ECD ( = 0.0175) and CCT ( = 0.0125) ( < 0.05). : The Nidek CEM-530 and the Tomey EM-4000 both produced reliable and reproducible results in terms of ECD and CCT. The absolute measurements were statistically significantly different for CCT and ECD for both devices; the Nidek produces slightly higher values.