Publication
Posterior Fossa Approaches Using the Leksell Vantage Frame with a Virtual Planning Approach in a Series of 10 Patients-Feasibility, Accuracy, and Pitfalls.
Journal Paper/Review - Nov 24, 2022
Krüger Marie Therese, Terrapon Alexis Paul Romain, Hoyningen Alexander, Kim Chan-Hi Olaf, Lauber Arno, Bozinov Oliver
Units
PubMed
Doi
Contact
Citation
Type
Journal
Publication Date
Issn Print
Brief description/objective
The open-face design of the Leksell Vantage frame provides many advantages. However, its more rigid, contoured design offers less flexibility than other frames. This is especially true for posterior fossa approaches. This study explores whether these limitations can be overcome by tailored frame placement using a virtual planning approach. The posterior fossa was accessed in ten patients using the Leksell Vantage frame. Frame placement was planned with the Brainlab Elements software, including a phantom-based (virtual) pre-operative planning approach. A biopsy was performed in all patients; in four, additional laser ablation surgery was performed. The accuracy of virtual frame placement was compared to actual frame placement. The posterior approach was feasible in all patients. In one case, the trajectory had to be adjusted; in another, the trajectory was switched from a right- to a left-sided approach. Both cases showed large deviations from the initially planned frame placement. A histopathological diagnosis was achieved in all patients. The new Leksell Vantage frame can be used to safely target the posterior fossa with a high diagnostic success rate and accuracy. Frame placement needs to be well-planned and executed. This can be facilitated using specific software solutions as demonstrated.