Chest X-ray Dose Equivalent Low-dose CT with Tin Filtration: Potential Role for the Assessment of Pectus Excavatum
Journal Paper/Review - Aug 28, 2019
Messerli-Odermatt Olivia, Serrallach Bettina, Gubser Manuel, Leschka Sebastian, Bauer Ralf, Dubois Josée, Alkadhi Hatem, Wildermuth Simon, Wälti Stephan
Units
PubMed
Doi
Citation
Type
Journal
Publication Date
Issn Electronic
Pages
Brief description/objective
RATIONALE AND OBJECTIVES
To determine the value of chest CT with tin filtration applying a dose equivalent to chest x-ray for the assessment of the Haller index for evaluation of pectus excavatum.
MATERIALS AND METHODS
Two hundred seventy-two patients from a prospective single center study were included and underwent a clinical standard dose chest CT (effective dose 1.8 ± 0.7 mSv) followed by a low-dose CT (0.13 ± 0.01 mSv) in the same session. Two blinded readers independently evaluated all data sets. Image quality for bony chest wall assessment was noted. Radiologists further assessed (a) transverse thoracic diameter, (b) anteroposterior thoracic diameter, and calculated (c) Haller index by dividing transverse diameter by anteroposterior diameter. The agreement of both readers in standard dose and low-dose CT was assessed using Lin's concordance correlation coefficient (p).
RESULTS
Subjective image quality was lower for low dose compared to standard dose CT images by both readers (p < 0.001). In total, 99% (n = 540) of low-dose CT scans were rated as diagnostic for bony chest wall assessment by both readers. There was a high agreement for assessment of transverse diameter, anteroposterior diameter and Haller index comparing both readers in standard dose and low-dose CT with p values indicating substantial agreement (i.e., 0.95> and ≤0.99) in 12/18 (67%) and almost perfect agreement (i.e., >0.99) in 6/18 (33%).
CONCLUSION
Our study suggests that low-dose CT with tin filtration applying a radiation dose equivalent to a plain chest X-ray is excellent for assessing the Haller index.