Publication
Tomotherapy dosimetry intercomparison 2016
Journal Paper/Review - Dec 1, 2016
Schiefer Johann, Buchauer Konrad, Heinze Simon
Units
Keywords
Link
Contact
Citation
Type
Journal
Publication Date
Pages
Publisher
Brief description/objective
It was the aim of the Tomotherapy dosimetry intercomparison of the year 2016 to check the absolute dosimetry. The cheese phantom setup, as established in the 2014 dosimetry intercomparison, was applied (Schiefer H, Buchauer K, Heinze S, Henke G, Plasswilm L, “Design and implementation of a "cheese" phantom-based Tomotherapy TLD dose intercomparison”, Strahlenther Onkol. 2015 Nov; 191(11):855-861. Epub 2015 Jun 19). All five Tomotherapy sites in Switzerland participated in the intercomparison. All dose point evaluations except two single TLD measurements led to a good result. All results are at minimum satisfactory.