Design and implementation of a "cheese" phantom-based Tomotherapy TLD dose intercomparison

Journal Paper/Review - Nov 1, 2015

Units
Keywords
Cylinder phantom; Dose intercomparison; TLD, thermoluminescent dosimetry; Tomotherapy; “Cheese” phantom
Doi
Link
Contact

Citation
Schiefer J. Design and implementation of a "cheese" phantom-based Tomotherapy TLD dose intercomparison. Strahlentherapie und Onkologie 2015; 191:855-861.
Type
Journal Paper/Review (English)
Journal
Strahlentherapie und Onkologie 2015; 191
Publication Date
Nov 1, 2015
Issn Print
0179-7158
Pages
855-861
Publisher
Springer (Heidelberg)
Brief description/objective

BACKGROUND:

The unique beam-delivery technique of Tomotherapy machines (Accuray Inc., Sunnyvale, Calif.) necessitates tailored quality assurance. This requirement also applies to external dose intercomparisons. Therefore, the aim of the 2014 SSRMP (Swiss Society of Radiobiology and Medical Physics) dosimetry intercomparison was to compare two set-ups with different phantoms.

MATERIALS AND METHODS:

A small cylindrical Perspex phantom, which is similar to the IROC phantom (Imaging and Radiation Oncology Core, Houston, Tex.), and the "cheese" phantom, which is provided by the Tomotherapy manufacturer to all institutions, were used. The standard calibration plans for the TomoHelical and TomoDirect irradiation techniques were applied. These plans are routinely used for dose output calibration in Tomotherapy institutions. We tested 20 Tomotherapy machines in Germany and Switzerland. The ratio of the measured (Dm) to the calculated (Dc) dose was assessed for both phantoms and irradiation techniques. The Dm/Dc distributions were determined to compare the suitability of the measurement set-ups investigated.

RESULTS:

The standard deviations of the TLD-measured (thermoluminescent dosimetry) Dm/Dc ratios for the "cheese" phantom were 1.9 % for the TomoHelical (19 measurements) and 1.2 % (11 measurements) for the TomoDirect irradiation techniques. The corresponding ratios for the Perspex phantom were 2.8 % (18 measurements) and 1.8 % (11 measurements).

CONCLUSION:

Compared with the Perspex phantom-based set-up, the "cheese" phantom-based set-up without individual planning was demonstrated to be more suitable for Tomotherapy dose checks. Future SSRMP dosimetry intercomparisons for Tomotherapy machines will therefore be based on the "cheese" phantom set-up.