Prevalence and morphology of coronary artery ectasia with dual-source CT coronary angiography

Journal Paper/Review - Dec 1, 2008

Units
PubMed
Doi
Contact

Citation
Leschka S, Stolzmann P, Scheffel H, Wildermuth S, Plass A, Genoni M, Marincek B, Alkadhi H. Prevalence and morphology of coronary artery ectasia with dual-source CT coronary angiography. Eur Radiol 2008; 18:2776-84.
Type
Journal Paper/Review (English)
Journal
Eur Radiol 2008; 18
Publication Date
Dec 1, 2008
Issn Electronic
1432-1084
Pages
2776-84
Brief description/objective

To assess the prevalence and morphological characteristics of coronary artery ectasia (CAE) with CT coronary angiography (CTCA) in comparison to conventional catheterangiography (CCA). Dual-source CTCA examinations from 677 consecutive patients (223 women; median age 57 years) were retrospectively evaluated by two blinded observers for the presence of CAE defined as a diameter enlargement > or = 1.5 times the diameter of adjacent normal coronary segments. Vessel diameters and contrast attenuation within and proximal to ectatic segments were measured. CCA was used to compare measurements obtained from CTCA with the coronary flow velocity by using the thrombolysis in myocardial infarction (TIMI) frame count. CTCA identified CAE in 20 of 677 (3%) patients. CCA was performed in ten of these patients. CAE diameter measurements with CTCA (10.0 +/- 5.4 mm) correlated significantly (r = 0.92, p < 0.001) with the CCA measurements (8.8 +/- 4.9 mm), but had higher diameters (levels of agreement: -1.0 to 3.4 mm). Contrast attenuation was significantly lower in the ectatic (343 +/- 63 HU) than in the proximal (394 +/- 60 HU) segments (p < 0.01). The attenuation difference significantly correlated with the CAE ratio (r = 0.67, p < 0.01) and the TIMI frame count (r = 0.58, p < 0.05). The prevalence of CAE in a population examined by CTCA is around 3%. Contrast attenuation measurements with CTCA correlate well with the flow alterations assessed with CCA.